Uncertainty Evaluation of Dynamic Pressure Calibrator by Monte Carlo Simulation |
Moon-Ki Kim |
몬테카를로 모사를 이용한 동압력 교정기 불확도 평가 |
김문기 |
국방과학연구소 |
|
Abstract |
This paper describes Monte Carlo Simulation(MCS) to assess the uncertainty of dynamic pressure calibrator and the expanded uncertainty results that were compared by GUM approximation and MCS. MCS uncertainties were computed using defining a domain of possible inputs, generating inputs randomly using probability distribution, performing a deterministic computation repeatedly and aggregating the results. It was revealed that the expanded uncertainty between GUM and MCS was different from each other. the expanded uncertainties were 0.5366%, 0.4856%, respectively. MCS is a suitable method for determining the uncertainty of simple and complex measurement systems. It should be more widely used and studied in measurement uncertainty calculations. |
Key Words:
MCS, Uncertainty, Dynamic Pressure Calibrator, GUM |
|